TRESAH

TRESA H

Compact, mobile effluent treatment for hematology analyzers

A compact, mobile decontamination unit for the highly concentrated effluent of hematology analyzers. A higher controlled dilution (1:10) precedes treatment by two 95 W UV-C lamps, and the unit fits next to or behind analyzers in space-limited laboratories.

Treatment capacity
12 L/h
UV-C lamps
2 × 95 W
Dilution level
1:10
Weight
37 kg
TRESA H stainless-steel mobile treatment unit on castors with control panel

Designed for

  • Hematology laboratories
  • Hospital laboratories
  • Private diagnostic laboratories
  • Central and reference laboratories
  • Medical facilities requiring controlled management of laboratory effluent

Key features

  • Integrated effluent holding tank
  • Gravity inlet with check valve
  • Automatic dilution water valve
  • Level monitoring and control
  • Dual UV-C treatment chambers
  • Integrated electrical and control system with status and fault monitoring
  • Stainless-steel construction, on castors
  • Accessible service components

Treatment process

Controlled dilution + UV-C. Level sensors and a solenoid valve control dilution before the effluent passes through dual UV-C treatment chambers.

  1. Collection

    Analyzer effluent flows by gravity through a check valve into the holding tank while air vents out.

  2. Controlled dilution

    Integrated level sensors and a solenoid valve add water to reach the required dilution ratio.

  3. UV-C activation

    The UV-C lamps switch on and preheat for optimal disinfection performance.

  4. Treatment & circulation

    A pump circulates the effluent through the dual UV-C chambers.

  5. Discharge

    Treated effluent is discharged to an approved drain or sewer connection, subject to local requirements.

Technical specifications

TRESA H technical specifications
ApplicationHematology analyzers
Analyzers per unitUp to 2
Treatment capacity12 L/h
Treatment technologyControlled dilution + UV-C
Dilution level1:10
UV-C lamps2 × 95 W
Lamp lifetime9,000 h
Power supply230 V
Weight37 kg
Dimensions (H × L × W)624 × 726 × 360 mm, on castors
ConstructionStainless steel

Values from the current TRESA H and TRESA range brochures. Specifications may change; please confirm with our team for your configuration.

Installation and utilities

Analyzerraw effluent
TRESA Htreatment on site
Drainapproved discharge point

Required connections

  • Analyzer effluent connection
  • Water supply for controlled dilution
  • Electrical power supply (230 V)
  • Treated-effluent discharge connection

Discharge of treated effluent must comply with the requirements that apply at your site. Local requirements and application conditions should be verified before installation.

TRESA H in the field and in production

Brochure

Documents are sent on request. Use the button to tell us which product you are interested in and we will reply with the relevant files.

TRESA H brochure

Specifications, process flow, system configuration and installation requirements.

PDF, 2 pages. Available on request.

Request product information

TRESA range brochure

Tri-fold brochure with a side-by-side comparison of all five TRESA systems, plus treatment overview and company details.

PDF, 2 pages. Available on request.

Request product information

TRESA H questions

How is TRESA H different from TRESA CI?

TRESA H is set up for hematology effluent, which is more concentrated, so it uses a higher 1:10 dilution and treats up to 12 L/h.

How many analyzers can TRESA H serve?

Up to two hematology analyzers.

What does TRESA H need on site?

A connection to the analyzer’s effluent line, a water supply for dilution, a 230 V supply and a drainage point.

Further reading

Knowledge Center

Ask about TRESA H for your laboratory

Tell us which analyzers you run and roughly how much effluent they produce. That helps our team advise which TRESA configuration may suit your site.