TRESA CIH brochure
Specifications, process flow, system configuration and installation requirements.
Request product informationFiltration and UV-C treatment in one cabinet for multi-analyzer effluent
A compact decontamination unit for the highly concentrated and non-concentrated effluent of biochemistry, immunoassay and hematology analyzers. Active-carbon filtration reduces chemical reagents and three UV-C lamps inactivate pathogens, treating up to 120 L/h.
Filtration + UV-C. Filtration reduces selected chemical contaminants before UV-C treatment provides microbial inactivation.
Concentrated and non-concentrated effluent from the analyzer flows into the collecting tank.
Active-carbon filtration (10 µm) removes most of the chemical reagents in the effluent.
Filtered effluent collects in a separate treatment tank, ready for UV-C processing.
Closed-loop circulation through three 254 nm UV-C lamps inactivates microorganisms.
Treated effluent is discharged to an approved drain or sewer connection, subject to local requirements.
| Application | Biochemistry, immunoassay and hematology analyzers |
|---|---|
| Analyzers per unit | Up to 3 |
| Treatment capacity | 120 L/h |
| Treatment technology | Active-carbon filtration + UV-C |
| Carbon filtration | 10 µm |
| UV-C lamps | 3 × 95 W, 254 nm |
| Lamp lifetime | 9,000 h |
| Control | PLC with touch display |
| Power supply | 230 V |
| Weight | 50 kg |
| Dimensions (H × L × W) | 700 × 800 × 600 mm |
| Maintenance | Filter cartridge every quarter; service once a year |
Values from the current TRESA CIH and TRESA range brochures. Specifications may change; please confirm with our team for your configuration.
Discharge of treated effluent must comply with the requirements that apply at your site. Local requirements and application conditions should be verified before installation.

Documents are sent on request. Use the button to tell us which product you are interested in and we will reply with the relevant files.
Specifications, process flow, system configuration and installation requirements.
Request product informationTri-fold brochure with a side-by-side comparison of all five TRESA systems, plus treatment overview and company details.
Request product informationUp to three biochemistry, immunoassay or hematology analyzers, treating both concentrated and non-concentrated effluent.
Active-carbon filtration (10 µm) reduces chemical reagents, then closed-loop circulation through three 95 W, 254 nm UV-C lamps inactivates microorganisms.
The filter cartridge is changed every quarter and the unit is serviced once a year.
Where liquid waste from IVD analyzers comes from, why it needs managing, and the options for treating it on site or collecting it for off-site disposal.
TRESA CITRESA CIHTRESA TL
How collecting analyzer effluent for off-site disposal compares with treating it on site, step by step.
TRESA CITRESA CIHTRESA TL
How 254 nm UV-C inactivates microorganisms in analyzer effluent, and why dilution or filtration comes first.
TRESA CITRESA HTRESA CIHTRESA CRD
Tell us which analyzers you run and roughly how much effluent they produce. That helps our team advise which TRESA configuration may suit your site.